GE Healthcare IN Cell Analyzer 2000 Cellular Imaging System: Partially Tested
Due to not having a copy of the acquisition software, our testing was limited to simply powering the instrument and internal computer on.
When we initially received the system, the IN Cell 2000 was switching on and off constantly. Our technician found that the computer, which controls when the analyzer powers on, had a faulty power adapter. He replaced the power adapter and rewired the power supply*, then replaced the PC's CMOS battery.
The computer was directly tested and found to be fully functional.
*Our technician rewired the power supply to fix a mistake made by the previous owner. The previous owner tried to fix the faulty power adapter by rewiring it to a separate supply but connected the negative and positive wires to the same terminal. There is a chance this damaged the unit, but we have seen no indications of any such damage.
In Excellent Condition
Filters Galore
This IN Cell 2000 comes with numerous filters, including:
- 34-100404-000 DV ET500/20 Y
- 34-100402-000 DV ET430/24 Y
- 34-100412-000 DV ET555/25 Y
- 34-100414-000 DV ET645/30 Y
- 34-100503-000 ET549/14 Y
- 34-100501-000 ET578/34 EX
- 34-100410-000 DV ET490/20 Y
- 34-100408-000DV360/40X
Along with the filters, this system comes packaged with:
- Ethernet cable
- Power cable
- Manuals
- Replacement amplifier
- Replacement wavetuner
- x2 copies of analysis software
- Computer documentation
- New door mechanism boxer
Note: There is a laser included as well, but it is likely defective.
Note: As mentioned above, this system does not include the acquisition software.
Specifications
Manufacturer
|
GE Healthcare Life Sciences
|
Model
|
IN Cell Analyzer 2000
|
Product No.
|
52-851714-001
|
Serial No.
|
B20086
|
Mfg. Date
|
September 2010
|
Electrical Requirements
|
100-120V, 50-60Hz, 1400VA
|
Dimensions (W x D x H)
|
44.5" x 26" x 24.5" (113 x 66 x 24.5 cm)
|