- Thermo Scientific Wellwash Versa Microplate Washer
- 2x8 Cell Wash Head
- x3 Bottles with Liquid Level Sensors
- Waste Bottle with Liquid Level Sensor
- Power Cable
- 120-Day Warranty
After its initial disinfection and inspection, our technicians tested the Thermo Scientific Wellwash Versa microplate washer with the following results:
- Washed several plates: PASSED
- System recognized when reagents were low: PASSED
- System recognized when waste bottle was full: PASSED
Manufacturer
|
Thermo Fisher Scientific
|
Model
|
Wellwash Versa
|
Catalog/Reference No.
|
5165010
|
Serial No.
|
888-1309B
|
Mfg. Date
|
August 2012
|
Type
|
888
|
Plate types
|
96- and 384-well plates Maximum height 16 mm
|
Wash volume
|
50–1000 ul in 50 ul increments (96-well plate) 20–120 ul in 10 ul increments (384-well plate)
|
Wash cycles
|
1-10
|
Residual volume
|
< 2.0 ul per well with sweep aspirate mode and with high aspirate speed at room temperature (flat-bottom 96-well plate)
|
Aspirate height
|
Adjustable, 0-14 mm in 0.1 mm increments
|
Dispense volume
|
50–400 ul in 50 ul increments (96-well plate) 20–120 ul in 10 ul increments (384-well plate)
|
Dispense accuracy
|
< 5% @ 300 ul at room temperature, with nx8 or nx12 wash head < 5% @ 250 ul, with 2x8 cell wash head < 5% @ 100 ul, with 1x16 wash head
|
Dispense precision
|
< 3% (CV) @ 300 ul, with nx8, nx12, and 1x16 wash heads
|
Soak time
|
1 s – 60 min in increments of 1 s/min
|
Rinse volume
|
5–100 ml in 5 ml increments
|
Prime volume
|
5–100 ml in 5 ml increments
|
Memory
|
At least 100 protocols
|
Linear shaker
|
Amplitude 2.5 mm; 5 Hz Amplitude 1.5 mm; 10 Hz Amplitude 1 mm; 15 Hz
|
Power Requirements
|
100-240 VAC, 50/60 Hz, 80 VA
|
Dimensions (W x D x H)
|
13 x 14.75 x 9.5 in
|