Applied Biosystems ABI 7900HT PCR DNA Sequence Detection Amplification System

For Parts or Repair. Error 3008.
$949.99
AdID: 14389

Availability: In stock

Details

For Parts or Repair - 7900HT Real-Time PCR System

This Applied Biosystems 7900HT was evaluated here in the NLS tech shop.


Incredibly clean and with nearly zero hours on the laser’s usage meter, this
unit could prove to be an excellent unit with lots of life left if properly re-
paired by an individual with the right expertise.


The issue with this 7900HT is that its tray motor is not calibrated.


We can connect to the 7900HT via the SDS software. Upon pressing the open/close
button, the door opens and the tray arm oscillates but not extends. The door
refuses close upon subsequent button presses.


The error log reports “3008: Tray Motor not Calibrated”.


We believe that this unit could make a great deal for someone with the knowledge and resources to fix it.


This instrument comes with a 384-well microplate block and a power cord.


Manufacturer Specifications

Manufacturer

Applied Biosystems

Model

7900HT

Serial No.

279000546

Part No.

4330966

Mfg. Date

November 2005

Dimensions (W x D x H)

21" x 34" x 24.75" (53.3 x 86.4 x 62.9 cm)